146 Chapter 12
Occlusion detection
The detection of an occlusion will trigger a “no delivery alarm.” Up
to 3.0 units of “missed” insulin are required to trigger the occlu-
sion alarm. If the occlusion is released, this insulin will be deliv-
ered. The table below demonstrates occlusion detections based on
U100 insulin. Three different rates of insulin delivery are shown.
Each with a minimum, typical and maximum time to trigger a “no
delivery alarm.”
Personal basal delivery patterns
Allows programming of up to three different basal patterns:
Standard, A and B.
Power supply
The pump is powered by a standard 1.5 V AAA alkaline battery.
(Energizer recommended for best results.)
The remote is powered by an A23 alkaline battery. Battery life,
approximately 6 months.
Rate
Minimum
Time
Typical Time
Maximum
Time
bolus delivery
(1.5u/minute)
10 seconds 25 seconds 2 minutes
1.0u/h 30 minutes 75 minutes 3 hours
0.1u/h 5 hours 12.5 hours 30 hours
Vedere la pagina 157