102 Chapter 7
8. Disconnect pump from your body. Remove reservoir. Press ACT
to rewind.
9. The pump will now rewind.
10. When the pump has finished rewinding the INSULIN TYPE
screen will appear. Use the and buttons to select your
reservoir type/insulin concentration. Press ACT.
11. If you do not press ACT immediately the screen will automati-
cally change to the INSULIN TYPE screen.
WARNING: Proper selection of insulin concentration is essential for
accurate insulin delivery and occlusion detection.
MiniMed
REWINDING
Wait for Notification
MiniMed
User-fill
INSULIN TYPE
User-fill
U50u
U100u